Robust Flow Control in a Mechatronic Test Lung with Spontaneous Breathing
IEEE International Conference on Advanced Intelligent Mechatronics (AIM), pp. 873–878
Abstract
To systematically evaluate mechanical ventilators and their automation concepts, test lungs can be used in hardware-in-the-loop (HiL) simulations. However, most test lungs do not simulate lungs with spontaneous breathing. This study presents a mechatronic test lung that consists of a bellow and a voice coil actuator. After modeling and parameter estimation, the model was used for the synthesis of a robust controller. The whole system was validated with a reference lung model and recorded measurement data of volume flow. The model identification resulted in a root-mean-square error (RMSE) of $\mathbf{0.96} \mathbf{~ m m}$. The controller achieved a maximum RMSE of 0.062 liters per second and a relative compliance error of $5.88 \%$. The test lung has the potential to enable HiL simulations of mechanical ventilation with spontaneous breathing.
Authors 6
-
Affiliation as printed
RWTH Aachen University,Chair for Medical Information Technology (MedIT),Aachen,Germany
-
Farid Hamidov Aachen
Affiliation as printed
RWTH Aachen University,MedIT,Aachen,Germany
-
Patrick Borchers Aachen
Affiliation as printed
RWTH Aachen University,MedIT,Aachen,Germany
-
Philip von Platen Aachen
Affiliation as printed
RWTH Aachen University,MedIT,Aachen,Germany
-
Steffen K. Leonhardt Aachen
Affiliation as printed
RWTH Aachen University,MedIT,Aachen,Germany
-
Marian Walter Aachen
Affiliation as printed
RWTH Aachen University,MedIT,Aachen,Germany
Cited by 1 stored of 1
1 result
No patents citing this paper on Lens.org (checked 2026-10-06).
References 13
-
W6637287672details pending0citations
13 results