Blood trauma in veno-venous extracorporeal membrane oxygenation: low pump pressures and low circuit resistance matter
Critical Care, vol. 28, pp. 330
Abstract
BACKGROUND: Veno-venous extracorporeal membrane oxygenation (VV ECMO) has become standard of care in patients with the most severe forms of acute respiratory distress syndrome. However, hemolysis and bleeding are one of the most frequent side effects, affecting mortality. Despite the widespread use of VV ECMO, current protocols lack detailed, in-vivo data-based recommendations for safe ECMO pump operating conditions. This study aims to comprehensively analyze the impact of VV ECMO pump operating conditions on hemolysis by combining in-silico modeling and clinical data analysis. METHODS: We combined data from 580 patients treated with VV ECMO in conjunction with numerical predictions of hemolysis using computational fluid dynamics and reduced order modeling of the Rotaflow (Getinge) and DP3 (Xenios) pumps. Blood trauma parameters across 94,779 pump operating points were associated with numerical predictions of shear induced hemolysis. RESULTS: Minimal hemolysis was observed at low pump pressures and low circuit resistance across all flow rates, whereas high pump pressures and circuit resistance consistently precipitated substantial hemolysis, irrespective of flow rate. However, the lower the flow rate, the more pronounced the influence of circuit resistance on hemolysis became. Numerical models validated against clinical data demonstrated a strong association (Spearman's r = 0.8) between simulated and observed hemolysis, irrespective of the pump type. CONCLUSIONS: Integrating in-silico predictions with clinical data provided a novel approach in understanding and potentially reducing blood trauma in VV ECMO. This study further demonstrated that a key factor in lowering side effects of ECMO support is the maintenance of low circuit resistance, including oxygenators with the lowest possible resistance, the shortest feasible circuit tubing, and cannulae with an optimal diameter.
Authors 6
-
Christopher Blum corresponding Aachen Medical Faculty Institute of Applied Medical Engineering Department of Cardiovascular Engineering
Affiliation as printed
Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany. blum@ame.rwth-aachen.de
Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany
-
Micha Landoll Aachen Medical Faculty Institute of Applied Medical Engineering Department of Cardiovascular Engineering
Affiliation as printed
Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany
Department of Pneumology and Critical Care Medicine, ARDS and ECMO Centre, Cologne-Merheim Hospital, Cologne, Germany
-
Affiliation as printed
Department of Pneumology and Critical Care Medicine, ARDS and ECMO Centre, Cologne-Merheim Hospital, Cologne, Germany
-
Ulrich Steinseifer Aachen Medical Faculty Institute of Applied Medical Engineering Department of Cardiovascular Engineering
Affiliation as printed
Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany
-
Michael Neidlin Aachen Medical Faculty Institute of Applied Medical Engineering Department of Cardiovascular Engineering
Affiliation as printed
Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany
-
Affiliation as printed
Department of Pneumology and Critical Care Medicine, ARDS and ECMO Centre, Cologne-Merheim Hospital, Cologne, Germany
University Witten/Herdecke, Witten, Germany
Cited by 17 stored of 17
17 results
No patents citing this paper on Lens.org (checked 2026-10-06).
References 26
-
W1742244419details pending0citations
-
W2041726928details pending0citations
-
W2127354627details pending0citations
-
W2130444042details pending0citations