Validation of aortic valve pressure gradient quantification using semi-automated 4D flow CMR pipeline
BMC Research Notes, vol. 15, pp. 151
Abstract
OBJECTIVE: Doppler echocardiographic aortic valve peak velocity and peak pressure gradient assessment across the aortic valve (AV) is the mainstay for diagnosing aortic stenosis. Four-dimensional flow cardiovascular magnetic resonance (4D flow CMR) is emerging as a valuable diagnostic tool for estimating the peak pressure drop across the aortic valve, but assessment remains cumbersome. We aimed to validate a novel semi-automated pipeline 4D flow CMR method of assessing peak aortic value pressure gradient (AVPG) using the commercially available software solution, CAAS MR Solutions, against invasive angiographic methods. RESULTS: We enrolled 11 patients with severe AS on echocardiography from the EurValve programme. All patients had pre-intervention doppler echocardiography, invasive cardiac catheterisation with peak pressure drop assessment across the AV and 4D flow CMR. The peak AVPG was 51.9 ± 35.2 mmHg using the invasive pressure drop method and 52.2 ± 29.2 mmHg for the 4D flow CMR method (semi-automated pipeline), with good correlation between the two methods (r = 0.70, p = 0.017). Assessment of AVPG by 4D flow CMR using the novel semi-automated pipeline method shows excellent agreement to invasive assessment when compared to doppler-based methods and advocate for its use as complementary to echocardiography.
Authors 12
-
Affiliation as printed
Norwich Medical School, University of East Anglia, Norwich, S10 2RX, UK
-
Affiliation as printed
Norwich Medical School, University of East Anglia, Norwich, S10 2RX, UK
-
Affiliation as printed
Pie Medical Imaging, Maastricht, The Netherlands
-
Affiliation as printed
Pie Medical Imaging, Maastricht, The Netherlands
-
National Heart Centre Singapore
Affiliation as printed
National Heart Centre, Singapore, Singapore
-
Jos J.M. Westenberg Aachen
Leiden University Medical Center
Affiliation as printed
Leids Universitair Medisch Centrum, Leiden, The Netherlands
-
Affiliation as printed
University of Sheffield, Sheffield, UK
-
Affiliation as printed
University of Sheffield, Sheffield, UK
-
Affiliation as printed
University of Sheffield, Sheffield, UK
-
Affiliation as printed
Norwich Medical School, University of East Anglia, Norwich, S10 2RX, UK
-
Affiliation as printed
Norwich Medical School, University of East Anglia, Norwich, S10 2RX, UK
-
Pankaj Garg corresponding
Affiliation as printed
Norwich Medical School, University of East Anglia, Norwich, S10 2RX, UK. p.garg@uea.ac.uk
Norwich Medical School, University of East Anglia, Norwich, S10 2RX, UK
Cited by 7 stored of 7
7 results
No patents citing this paper on Lens.org (checked 2026-10-11).
References 13
13 results