Flow profile generation for a left ventricular assist device using iterative learning control

Current Directions in Biomedical Engineering, vol. 7, pp. 279–282

Abstract

Abstract Nowadays, rotary blood pumps for the treatment of end-stage heart failure patients are usually fixed-speed controlled. Therefore, without sufficient heart activity, patients may reveal reduced blood flow pulsatility. However, there is evidence that pulsatile flow reduces adverse events. Furthermore, optimized pump flow profiles could be used to achieve certain control objectives. Therefore, a pump flow model is identified and subsequently used to design a control system with an iterative learning controller to generate desired pump flow patterns. Either a flow sensor is required for direct measurement or the flow can be estimated from pressure sensor readings using the proposed pump flow model. For comparison, a PID controller with disturbance feedforward control is also designed. Furthermore, the robustness against respiratory sinus arrhythmia is analyzed.

Authors 5

  1. RWTH Aachen University

    Affiliation as printed

    RWTH Aachen University, Pauwelsstraße 20, 52074 Aachen , Germany

  2. Marian Walter Aachen

    RWTH Aachen University

    Affiliation as printed

    RWTH Aachen University, Pauwelsstraße 20, 52074 Aachen , Germany

  3. RWTH Aachen University

    Affiliation as printed

    RWTH Aachen University, Pauwelsstraße 20, 52074 Aachen , Germany

  4. National Research University of Electronic Technology

    Affiliation as printed

    National Research University of Electronic Technology, Moscow , Russia

  5. National Research University of Electronic Technology

    Affiliation as printed

    National Research University of Electronic Technology, Moscow , Russia

Cited by 4 stored of 4

4 results

No patents citing this paper on Lens.org (checked 2026-10-06).

References 8

8 results