Local SAR compression with overestimation control to reduce maximum relative SAR overestimation and improve multi-channel RF array performance

Magnetic Resonance Materials in Physics Biology and Medicine, vol. 34, pp. 153–163

Abstract

OBJECTIVE: In local SAR compression algorithms, the overestimation is generally not linearly dependent on actual local SAR. This can lead to large relative overestimation at low actual SAR values, unnecessarily constraining transmit array performance. METHOD: Two strategies are proposed to reduce maximum relative overestimation for a given number of VOPs. The first strategy uses an overestimation matrix that roughly approximates actual local SAR; the second strategy uses a small set of pre-calculated VOPs as the overestimation term for the compression. RESULT: Comparison with a previous method shows that for a given maximum relative overestimation the number of VOPs can be reduced by around 20% at the cost of a higher absolute overestimation at high actual local SAR values. CONCLUSION: The proposed strategies outperform a previously published strategy and can improve the SAR compression where maximum relative overestimation constrains the performance of parallel transmission.

Authors 5

  1. Stephan Orzada corresponding

    Erwin L. Hahn Institute for Magnetic Resonance Imaging · Essen University Hospital · University of Duisburg-Essen

    Affiliation as printed

    Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Kokereiallee 7, 45141, Essen, Germany. Stephan.orzada@uni-due.de

    High-Field and Hybrid MR Imaging, University Hospital Essen, 45147, Essen, Germany. Stephan.orzada@uni-due.de

    Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Kokereiallee 7, 45141, Essen, Germany

  2. German Cancer Research Center

    Affiliation as printed

    Medical Physics in Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany

  3. FH Aachen

    Affiliation as printed

    Electromagnetic Theory and Applied Mathematics, Faculty of Electrical Engineering and Information Technology, FH Aachen, University of Applied Sciences, 52066, Aachen, Germany

  4. Erwin L. Hahn Institute for Magnetic Resonance Imaging · German Cancer Research Center · Heidelberg University · University of Duisburg-Essen

    Affiliation as printed

    Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Kokereiallee 7, 45141, Essen, Germany

    Faculty of Medicine, University of Heidelberg, Im Neuenheimer Feld 672, 69120, Heidelberg, Germany

    Faculty of Physics and Astronomy, University of Heidelberg, Im Neuenheimer Feld 226, 69120, Heidelberg, Germany

    Medical Physics in Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany

  5. Erwin L. Hahn Institute for Magnetic Resonance Imaging · Essen University Hospital · University of Duisburg-Essen

    Affiliation as printed

    Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Kokereiallee 7, 45141, Essen, Germany

    High-Field and Hybrid MR Imaging, University Hospital Essen, 45147, Essen, Germany

Cited by 11 stored of 11

11 results

No patents citing this paper on Lens.org (checked 2026-10-06).

References 26