HeartWare HVAD Flow Estimator Accuracy for Left and Right Ventricular Support
ASAIO Journal, vol. 67, pp. 416–422
Abstract
This study investigated the accuracy of the HeartWare HVAD flow estimator for left ventricular assist device (LVAD) support and biventricular assist device (BiVAD) support for modes of reduced speed (BiVAD-RS) and banded outflow (BiVAD-B). The HVAD flow estimator was evaluated in a mock circulatory loop under changes in systemic and pulmonary vascular resistance, heart rate, central venous pressure, and simulated hematocrit (correlated to viscosity). A difference was found between mean estimated and mean measured flow for LVAD (0.1 ± 0.3 L/min), BiVAD-RS (-0.1 ± 0.2 L/min), and BiVAD-B (0 ± 0.2 L/min). Analysis of the flow waveform pulsatility showed good correlation for LVAD (r2 = 0.98) with a modest spread in error (0.7 ± 0.1 L/min), while BiVAD-RS and BiVAD-B showed similar spread in error (0.7 ± 0.3 and 0.7 ± 0.2 L/min, respectively), with much lower correlation (r2 = 0.85 and r2 = 0.60, respectively). This study demonstrated that the mean flow error of the HVAD flow estimator is similar when the device is used in LVAD, BiVAD-RS, or BiVAD-B configuration. However, the instantaneous flow waveform should be interpreted with caution, particularly in the cases of BiVAD support.
Authors 7
-
Andrew F. Stephens corresponding
Monash University · Baker Heart and Diabetes Institute
Affiliation as printed
Cardiorespiratory Engineering and Technology Laboratory, Baker Heart and Diabetes Institute, Melbourne, VIC, Australia
Department of Mechanical and Aerospace Engineering, Monash University, Melbourne, VIC, Australia
-
Griffith University · Prince Charles Hospital
Affiliation as printed
Innovative Cardiovascular Engineering and Technology Laboratory, Critical Care Research Group, The Prince Charles Hospital, Chermside, QLD, Australia
School of Engineering and Built Environment, Griffith University, Brisbane, QLD, Australia
-
The University of Queensland · Prince Charles Hospital
Affiliation as printed
Innovative Cardiovascular Engineering and Technology Laboratory, Critical Care Research Group, The Prince Charles Hospital, Chermside, QLD, Australia
School of Medicine, University of Queensland, Brisbane, QLD, Australia
-
The University of Queensland · Prince Charles Hospital
Affiliation as printed
Innovative Cardiovascular Engineering and Technology Laboratory, Critical Care Research Group, The Prince Charles Hospital, Chermside, QLD, Australia
School of Medicine, University of Queensland, Brisbane, QLD, Australia
-
RWTH Aachen University · Monash University · Baker Heart and Diabetes Institute
Affiliation as printed
Cardiorespiratory Engineering and Technology Laboratory, Baker Heart and Diabetes Institute, Melbourne, VIC, Australia
Cardiovascular Engineering, Rheinisch-Westfälische Technische Hochschule Aachen University, Aachen, Germany
Department of Mechanical and Aerospace Engineering, Monash University, Melbourne, VIC, Australia
-
Affiliation as printed
Innovative Cardiovascular Engineering and Technology Laboratory, Critical Care Research Group, The Prince Charles Hospital, Chermside, QLD, Australia
-
Shaun David Gregory corresponding
Monash University · Baker Heart and Diabetes Institute
Affiliation as printed
Cardiorespiratory Engineering and Technology Laboratory, Baker Heart and Diabetes Institute, Melbourne, VIC, Australia
Department of Mechanical and Aerospace Engineering, Monash University, Melbourne, VIC, Australia
Cited by 5 stored of 5
5 results
No patents citing this paper on Lens.org (checked 2026-10-06).