Applicator visualization using ultrashort echo time MRI for high-dose-rate endorectal brachytherapy
Brachytherapy, vol. 19, pp. 618–623
Abstract
Purpose The individual channels in an endorectal applicator for high-dose-rate endorectal brachytherapy are not visible on standard MRI sequences. The aim of this study was to test whether an ultrashort echo time (UTE) MRI sequence could be used to visualize the individual channels to enable MR-only treatment planning for rectal cancer. Methods and Materials We used a radial three-dimensional (3D) UTE pulse sequence and acquired images of phantoms and two patients with rectal cancer. We rigidly registered a UTE image and CT scan of an applicator phantom, based on the outline of the applicator. One observer compared channel positions on the UTE image and CT scan in five slices spaced 25 mm apart. To quantify geometric distortions, we scanned a commercial 3D geometric quality assurance phantom and calculated the difference between detected marker positions on the UTE image and corresponding marker positions on two 3D T 1 -weighted images with opposing readout directions. Results On the UTE images, there is sufficient contrast to discern the individual channels. The difference in channel positions on the UTE image compared with the CT was on average −0.1 ± 0.1 mm (left-right) and 0.1 ± 0.3 mm (anteroposterior). After rigid registration to the 3D T 1 -weighted sequences, the residual 95th percentile of the geometric distortion inside a 550-mm-diameter sphere was 1.0 mm (left-right), 0.9 mm (anteroposterior), and 0.9 mm (craniocaudal). Conclusions With a UTE sequence, the endorectal applicator and individual channels can be adequately visualized in both phantom and patients. The geometrical fidelity is within an acceptable range.
Authors 6
-
Leiden University Medical Center
Affiliation as printed
Department of Radiation Oncology, Leiden University Medical Center, Leiden, the Netherlands. Electronic address: r.p.j.van_den_ende@lumc.nl
Department of Radiation Oncology, Leiden University Medical Center, Leiden, the Netherlands
-
Ece Ercan Aachen C.J. Gorter Center for High Field MRI Department of Radiology, Leiden University Medical Center
Leiden University Medical Center
Affiliation as printed
C.J. Gorter Center for High Field MRI, Department of Radiology, Leiden University Medical Center, Leiden, the Netherlands
-
The Netherlands Cancer Institute
Affiliation as printed
Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, the Netherlands
-
Leiden University Medical Center
Affiliation as printed
Department of Radiation Oncology, Leiden University Medical Center, Leiden, the Netherlands
-
Leiden University Medical Center · The Netherlands Cancer Institute
Affiliation as printed
Department of Radiation Oncology, Leiden University Medical Center, Leiden, the Netherlands; Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, the Netherlands
Department of Radiation Oncology, Leiden University Medical Center, Leiden, the Netherlands
Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, the Netherlands
-
Leiden University Medical Center · The Netherlands Cancer Institute
Affiliation as printed
Department of Radiation Oncology, Leiden University Medical Center, Leiden, the Netherlands; Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, the Netherlands
Department of Radiation Oncology, Leiden University Medical Center, Leiden, the Netherlands
Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam, the Netherlands
Cited by 6 stored of 6
6 results
No patents citing this paper on Lens.org (checked 2026-10-11).
References 17
-
W6605497547details pending0citations
17 results