Usability Issues with Mechanical Ventilation Devices
Studies in health technology and informatics, vol. 270, pp. 1419–1420
Abstract
The user interface of a mechanical ventilator is safety critical, as use errors can lead to patient harm. A systematic review was conducted to identify published usability issues and contributing factors that can lead to use errors. The findings were grouped in an Ishikawa diagram. Many of the problems mentioned based on inconsistent labeling and manufacturer-specific naming of ventilation modes. In the studies, usability was often measured quantitatively and did not allow any conclusions to be drawn about concrete problems.
Authors 4
-
Affiliation as printed
Institute of Medical Informatics, Medical Faculty, RWTH Aachen University, Germany
-
Carl von Ossietzky Universität Oldenburg
Affiliation as printed
Division for Medical Informatics, Carl von Ossietzky University of Oldenburg, Germany
-
Affiliation as printed
Institute of Medical Informatics, Medical Faculty, RWTH Aachen University, Germany
-
Affiliation as printed
Institute of Medical Informatics, Medical Faculty, RWTH Aachen University, Germany
Cited by 2 stored of 2
2 results
No patents citing this paper on Lens.org (checked 2026-10-06).